Comparison of Central Corneal Thickness Measurements by Pentacam and Ultrasound Pachymetry in Normal Myopic Patients

Authors

  • Sherif Emerah Associate Professor of Ophthalmology, Ophthalmology department, Research Institute of Ophthalmology, Giza, Egypt

Keywords:

Corneal thickness, Pentacam, Ultrasound pachymetry

Abstract

Introduction: Precise measurement of corneal thickness is of crucial importance in corneal refractive procedures. The aim of this work is to compare the central corneal thickness measurements using Scheimpflug imaging (Pentacam) and contact US pachymetry in normal myopic patients. 

Methods: This retrospective cohort study was conducted on in 144 eyes of 72 normal myopic patients during 6 months. The study location was the Research Institute of Ophthalmology (Giza, Egypt), and Magrabi Eye Hospital (Dubai, United Arab Emirates). Central corneal thickness was measured with the two devices. Data were analyzed by descriptive statistics, independent-samples t-test, paired-samples t-test, Mann-Whitney U test and Pearson-Product Moment Correlation Coefficient.

Results: The mean age was 25.3±5.6 years (range 18 to 38 years). The median age for our patients is 24 years. The average measurements of CCT were 526.8±35.3 and 529.1±37.9 for the Pentacam pachymetry and US pachymetry values, respectively. Mean difference between both measurements was 2.3±2.6μm which is statistically insignificant (p=0.08). Bland Altman Plot shows border line agreement of the two methods US pachymetry and Pentacam in detection of central corneal thickness. 

Conclusion: The results of this study showed no significant differences in CCT readings between the measurements obtained by Pentacam and US pachymetry in normal myopic patients.

 

References

Kanski JJ. Clinical Ophthalmology A systematic approach. 5th ed. London: Butterworth-Heineman. 2003:

Cheng H, Bates AK, Wood L, McPherson K. Positive correlation of corneal thickness and endothelial cell

loss. Serial measurements after cataract surgery. Arch Ophthalmol. 1988; 106(7): 920-2. doi:

1001/archopht.1988.01060140066026. PMID: 3390055.

Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a

review and metaanalysis approach. Surv Ophthalmol. 2000; 44(5): 367-408. doi: 10.1016/S0039- 6257(00)00110-7. PMID: 10734239.

Auffarth GU, Wang L, Völcker HE. Keratoconus evaluation using the Orbscan topography system. Journal

of Cataract and Refractive Surgery. 2000; 26(2): 222-8. doi: 10.1016/S0886-3350(99)00355-7.

Amano S, Honda N, Amano Y, Yamagami S, Miyai T, Samejima T, et al. Comparison of central corneal

thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal

topography. Ophthalmology. 2006; 113(6): 937-41. doi: 10.1016/j.ophtha.2006.01.063. PMID: 16751036.

Ho T, Cheng AC, Rao SK, Lau S, Leung CK, Lam DS. Central corneal thickness measurements using

Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia. J

Cataract Refract Surg. 2007; 33(7): 1177-82. doi: 10.1016/j.jcrs.2007.03.028. PMID: 17586372.

Hashemi H, Mehravaran S. Central corneal thickness measurement with Pentacam, Orbscan II, and

ultrasound devices before and after laser refractive surgery for myopia. J Cataract Refract Surg. 2007;

(10): 1701-7. doi: 10.1016/j.jcrs.2007.05.040. PMID: 17889763.

Ciolino JB, Khachikian SS, Belin MW. Comparison of corneal thickness measurements by ultrasound and

Scheimpflug photography in eyes that have undergone laser in situ keratomileusis. Am J Ophthalmol.

; 145(1): 75-80. doi: 10.1016/j.ajo.2007.08.026. PMID: 17988643.

Sedaghat MR, Daneshvar R, Kargozar A, Derakhshan A, Daraei M. Comparison of central corneal

thickness measurement using ultrasonic pachymetry, rotating Scheimpflug camera, and scanning-slit

topography. Am J Ophthalmol. 2010; 150(6): 780-9. doi: 10.1016/j.ajo.2010.06.013. PMID: 20932508.

Al-Mezaine HS, Al-Amro SA, Kangave D, Al-Obeidan S, Al-Jubair KM. Comparison of central corneal

thickness measurements using Pentacam and ultrasonic pachymetry in post-LASIK eyes for myopia. Eur J

Ophthal. 2010; 20(5): 852-7. PMID: 20815096.

Tai LY, Khaw KW, Ng CM, Subrayan V. Central corneal thickness measurements with different imaging

devices and ultrasound pachymetry. Cornea. 2013; 32(6): 766-1. doi: 10.1097/ICO.0b013e318269938d.

PMID: 23095499.

Sadoughi MM, Einollahi B, Einollahi N, Rezaei J, Roshandel D, Feizi S. Measurement of Central Corneal

Thickness Using Ultrasound Pachymetry and Orbscan II in Normal Eyes. J Ophthalmic Vis Res. 2015;

(1): 4-9. doi: 10.4103/2008-322X.156084. PMID: 26005545, PMCID: PMC4424717.

Published

2022-03-07